Appendix 4: Meter factory default settings
Resetting the meter to factory default settings clears all calibration data and most other set up functions
you might have changed. The following settings will remain as you have set them:
• temperature unit of measure (°C or °F)
• The temperature offset calibration value
NOTE: Conductivity/TDS and pH data are cleared separately from each other.
• To clear pH data, see Sub group P4.0 on page 30
• To clear conductivity/TDS data, see Sub group P9.0 on page 37
Type
Parameter
Default
Remarks
pH parameters
P1.1
Viewing pH
– – –
No calibration data for pH 1.68
P1.2
calibration data
– – –
No calibration data for pH 4.01
P1.3
– – –
No calibration data for pH 7.00
P1.4
– – –
No calibration data for pH 10.00
P1.5
– – –
No calibration data for pH 12.45
P2.1
View electrode offset
0.00 mV
No offset adjustment
P2.2
View electrode slope
100.0%
No slope adjustment
P3.1
Ready indicator
Ready on
Ready indicator on; auto endpoint off
P3.2
# pH cal. points
3
3 pH calibration points available (of a 2-5 point range)
P3.3
NIST or USA buffer sets
USA
—
P3.4
°F or °C
no default
°F or °C remains as selected
P4.0
pH factory default
no
Retains your current settings
Conductivity/TDS parameters
P5.1
Viewing cond./TDS
– – –
No calibration data for Range 1 (0.00-19.99 µS / 0.00-9.99 ppm)
P5.2
calibration data
– – –
No calibration data for Range 2 (0.0-199.9 µS / 10.0-99.9 ppm)
P5.3
– – –
No calibration data for Range 3 (0-1999 µS / 100-999 ppm)
P5.4
– – –
No calibration data for Range 4 (0.00-19.99 mS / 1.00-9.99 ppt)
P5.5
– – –
No calibration data for Range 5 (0.0-199.9 mS / 10.0-199.9 ppt)
P6.1
Viewing cond./TDS
1.000
No offset for effective cell constant (0.00-19.99 µS / 0.00-9.99 ppm)
P6.2
effective cell constant
1.000
No offset for effective cell constant (0.0-199.9 µS / 10.0-99.9 ppm)
P6.3
1.000
No offset for effective cell constant (0-1999 µS / 100-999 ppm)
P6.4
1.000
No offset for effective cell constant (0.00-19.99 mS / 1.00-9.99 ppt)
P6.5
1.000
No offset for effective cell constant (0.0-199.9 mS / 10.0-199.9 ppt)
P7.1
Ready indicator
Ready on
Ready indicator on; auto endpoint off
P7.2
°F or °C
no default
°F or °C remains as selected
P7.3
ATC on or off
ATC on
—
P7.4
Set Cond-to-TDS factor
0.5
Adjustable from 0.4 to 1.0
P8.1
Temp. coefficient
2.1% per °C
Adjustable from 0 to 10% per *C
P8.2
Normalization temp
25°C
Adjustable from 15 to 30°C / 59 to 86°F
P9.0
cond./TDS factory default
no
Retains your current settings