10. Troubleshooting
Error Messages
The following table provides a guideline to enable diagnosis
of possible problems indicated by the messages generated
by the OAKTON pH/CON 510 meter. The table also provides
possible solutions to the problems encountered.
Error Message
Indicates
Cause
Corrective Action
Err
(In secondary display)
Wrong keypad input.
Button does not work
Release key. Select
in the current
valid buttons
operation mode.
depending on mode.
Electrode icon
(Indicator)
Calibration error
Buffer value does not match
Use fresh buffer or check
value displayed or electrode
electrode connection.
is disconnected or failing.
Possibly replace electrodes.
Ur or Or
Measured value
Electrode not connected
Make sure electrode is connected
(In primary display)
out of range
Meter not calibrated
Recalibrate meter
Electrode clogged, dirty or
Clean or replace electrode
broken
Measurement is under range
Treat samples to bring within
or over range
meter measuring range