Using manual ranging function:
conductivity or TDS
When shipped from the factory, your meter automatically
selects the range in which your readings appear. The manual
ranging function lets you select the specific range in which
you want to work.
From measurement mode:
1.
To select the desired measuring range, press the RANGE
key while in Conductivity (or TDS) measurement mode.
The first range will appear on the display. The “MEAS”
indicator blinks, and the “Range” number flashes briefly
in the lower display.
conductivity
TDS
r1 =
0.00-19.99 µS
0.00-9.99 ppm
r2 =
0.0-199.9 µS
10.0-99.9 ppm
r3 =
0-1999 µS
100-999 ppm
r4 =
0.00-19.99 mS
1.00-9.99 ppt
r5 =
0.0-199.9 mS
10.0-199.9 ppt
See Figure 24
2.
Press the RANGE key again (if needed) until desired
range is selected.
3.
To reselect the Auto-ranging function, repeatedly press
the RANGE key until the “MEAS” indicator appears with-
out blinking.
Notes
This meter will not let you manually select a range in which
the reading will be overrange.
The meter resets to the Auto-ranging function once it is
turned off. You will have reset the manual ranging function
each time you turn the meter off.