Selecting manual temperature compensation:
conductivity or TDS
For manual temperature compensation in conductivity or
TDS mode, you must:
1.
Deactivate the temperature element built into the
conductivity/TDS probe.
2.
Select a manual temperature compensation value
(see page 21).
You can deactivate the temperature element (select manual
temperature compensation) in Set Up Program P7.3.
From measurement mode
1.
Press the Mode key to select conductivity (or TDS)
measurement mode.
2.
Press the SET key to enter Set Up mode.
3.
Press the
▲
or
▼
keys to scroll through subgroups until
you view parameter P7.0.
See Figure 25
4.
Press the ENTER key three times to select parameter 7.3.
The upper display shows “ATC” and the lower display
shows “P7.3”.
See Figure 26
5.
Press the ENTER key again. The upper display shows
“ATC” and the lower display shows “YES” or “NO”.
6.
Press the
▲
and
▼
keys to select the Automatic
Temperature Compensation off (NO).
• YES = ATC on
• NO = ATC off (manual compensation)
See Figure 27
7.
Press the ENTER key to confirm selection and to return to
the subgroup menu. Press the CAL/MEAS key twice to
return to measurement mode.
Note that the ATC indicator no longer appears on the
display.
Go to page 21 to select a manual temperature
compensation value.