This mode lets you recall previous conductivity or TDS
calibration data, which lets you know at which points this
meter was previously calibrated. This is a “view only” mode.
From measurement mode:
1.
Press the Mode key to select conductivity (TDS) measure-
ment mode.
2.
Press the SET key to enter Set Up mode.
3.
Press the
▲
and
▼
keys to scroll through subgroups until
you view parameter P5.0.
See Figure 44
4.
Press the ENTER key repeatedly to view previous
calibration data.
conductivity
TDS
P5.1 (Range 1) =
0.00-19.99 µS
0.00-9.99 ppm
P5.2 (Range 2) =
0.0-199.9 µS
10.0-99.9 ppm
P5.3 (Range 3) =
0-1999 µS
100-999 ppm
P5.4 (Range 4)=
0.00-19.99 mS
1.00-9.99 ppt
P5.5 (Range 5)=
0.0-199.9 mS
10.0-199.9 ppt
See Figure 45
5.
When you have scrolled through all calibration data, you
will automatically return to the subgroup menu. Press the
CAL/MEAS key to return to measurement mode.
Notes
If there is no previous calibration data at a particular point,
the primary display will show “– – –”.
See Figure 46