25
P9.0: Reset to factory default (conductivity or TDS)
Reset conductivity data to factory default settings
mS
MEAS
SETUP
µ
S
ppt
ppm
P5.0: Viewing conductivity or TDS calibration data
P5.1 View conductivity/TDS calibration data R1
R1 = 0.00-19.99 µS / 0.00-9.99 ppm
P5.2 View conductivity/TDS calibration data R2
R2 = 0.0-199.9 µS / 10.0-99.9 ppm
P5.3 View conductivity/TDS calibration data R3
R3 = 0-1999 µS / 100-999 ppm
P5.4 View conductivity/TDS calibration data R4
R4 = 0.00-19.99 mS / 1.00-9.99 ppt
P5.5 View conductivity/TDS calibration data R5
R5 = 0.0-199.9 mS / 10.0-199.9 ppt
P6.0: Viewing conductivity or TDS probe data
P6.1 View effective cell constant R1
R1 = 0.00-19.99 µS / 0.00-9.99 ppm
P6.2 View effective cell constant R2
R2 = 0.0-199.9 µS / 10.0-99.9 ppm
P6.3 View effective cell constant R3
R3 = 0-1999 µS / 100-999 ppm
P6.4 View effective cell constant R4
R4 = 0.00-19.99 mS / 1.00-9.99 ppt
P6.5 View effective cell constant R5
R5 = 0.0-199.9 mS / 10.0-199.9 ppt
P7.0: Conductivity or TDS configuration
P7.1 Ready indicator on or off / auto endpoint on or off
P7.2 Select °F or °C
P7.3 Select Automatic or Manual Temperature
Compensation
P7.4 Set Conductivity-to-TDS conversion factor
P8.0: Temperature parameters
P8.1 Set temperature coefficient
P8.2 Set normalization temperature
mS
SETUP
µ
S
ppt
ppm
mS
SETUP
µ
S
ppt
ppm
mS
SETUP
µ
S
ppt
ppm
mS
F
SETUP
C
µ
S
ppt
ppm
Sub groups available from conductivity or TDS measurement mode:
troubleshooting
electrode
care
measurement
calibration
starting up
set up
mode
return of
items
warranty
accessories
specifications
appendixes