32
P6.0: Viewing conductivity (TDS) probe data
Program 6 has five "view only" options that let you check
your conductivity/TDS probe’s parameters for diagnostic
purposes. These options show you the effective cell
constant for each range. The cell constant is adjusted
according to your calibration.
From measurement mode:
1.
Press the Mode key to select conductivity (TDS)
measurement mode.
2
. Press the SET key to enter Set Up mode.
3.
Press the
▲
and
▼
keys to scroll through subgroups until
you view parameter P6.0.
See Figure 47
4.
Press the ENTER key repeatedly to view the effective cell
constant for each range.
conductivity
TDS
P6.1 (Range 1) =
0.00-19.99 µS
0.00-9.99 ppm
P6.2 (Range 2) =
0.0-199.9 µS
10.0-99.9 ppm
P6.3 (Range 3) =
0-1999 µS
100-999 ppm
P6.4 (Range 4)=
0.00-19.99 mS
1.00-9.99 ppt
P6.5 (Range 5)=
0.0-199.9 mS
10.0-199.9 ppt
See Figure 48
5.
When you have scrolled through all probe data, you will
automatically return to the subgroup menu. Press
the CAL/MEAS key to return to measurement mode.
Notes
Cell constants will degrade with time and usage. You can use
this feature to alert you to when your probe needs cleaning
or to when you need to replace your probe.