P8.0: Temperature Settings
P8.1 Selecting the temperature coefficient
The temperature coefficient is the amount of change in conduc-
tivity per degree of temperature. Entering the exact temperature
coefficient of your solution lets you accurately compensate tem-
perature for almost any solution. You can adjust 0.0 to 10.0%
per °C or °F. Meter default is 2.1% per °C or °F.
Determine your solution’s temperature coefficient using the for-
mula in App. 3 “Calculating Temperature Coefficients”, page 45.
From measurement mode
1.
Press Mode to select conductivity/TDS measurement mode.
2.
Press the SET key to enter Set Up mode.
3.
Press the
▲
and
▼
keys to scroll through subgroups until you
view parameter P8.0.
See Figure 56
4
. Press the ENTER key to select parameter 8.1. The display
shows “T.CO” on the upper display.
See Figure 57
5.
Press the ENTER key again.. The upper display shows the
temperature coefficient and the lower display shows “T.CO”.
See Figure 58
6.
Press the
▲
and
▼
keys to select the temperature coeffi-
cient of your solution.
7.
Press the ENTER key to confirm selection and to proceed to
step 5 of P.8.2. Press the CAL/MEAS key twice to return to
measurement mode.
P8.2 Selecting the normalization temperature
Your meter normalizes its conductivity (TDS) measurements to a
standard temperature. You can adjust the normalization temper-
ature from 15 to 30°C (59 to 86°F). Meter default is 25°C (77°F).
From measurement mode
1.
Press Mode to select conductivity/TDS measurement mode.
2.
Press the SET key to enter Set Up mode.
3.
Press the
▲
and
▼
keys to scroll through subgroups until you
view parameter P8.0.
See Figure 56
4.
Press the ENTER key three times to select parameter 8.2.
The display shows “t.nr” on the upper display.
See Figure 59
5.
Press the ENTER key again. The upper display shows the nor-
malization temperature and the lower display shows “t.nr”.
See Figure 60
6.
Press the
▲
and
▼
keys to select normalization temperature.
7.
Press the ENTER key to confirm selection and to return to
the subgroup menu. Press the CAL/MEAS key to return to
measurement mode.
F
SETUP
C
mS
µ
S
ppt
ppm
Figure 56
F
SETUP
C
Figure 57
SETUP
Figure 58
F
SETUP
C
Figure 59
SETUP
C
Figure 60
troubleshooting
electrode
care
measurement
calibration
starting up
set up
mode
return of
items
warranty
accessories
specifications
appendixes