41
Operation of BMH Series
English
5 – 2 – 2 Bone Saw Attachment 200 (P200-SAG)
Mounting the Irrigation Nozzle (Optional)
_ Mounting
1
Insert the Irrigation Nozzle
Ⓐ
onto the Bur Guard
Ⓑ
of the
Attachment, making sure that the correct end of the Irrigation
Nozzle goes in.
2
Insert the Irrigation Tube
Ⓒ
into the Irrigation Nozzle
Ⓐ
.
Ⓐ
Ⓑ
Ⓒ
_ Removing
Perform steps 1 and 2 of the mounting procedure in the reverse
order.
* For the related connection and the setting on the Control Unit, refer to OM-SE0021E Primado2 OPERATION MANUAL Control Unit and Foot
Controls.
Mounting and Removing the Cutting Accessories (Optional)
_ Mounting
1
While depressing the Fitting/Removal Button
Ⓐ
, fit the
Alignment Holes of a Cutting Accessory
Ⓑ
into the convex part
of the Attachment
Ⓒ
.
Ⓐ
Ⓒ
Ⓑ
Ⓐ
2
Release the Fitting/Removal Button
Ⓐ
to set the Cutting
Accessory in place.
Ⓐ
Ⓐ
3
By twisting the Adjusting Nut
Ⓓ
clockwise and
counterclockwise, set the Cutting Accessory in place properly.
Ⓓ
Ⓓ
4
After mounting the Cutting Accessory, pull it lightly to confirm
that it is securely connected to the Attachment.
Ⓔ
* If the Cutting Accessory is not set in place properly, repeat the
procedure from 1 - 3 to set it in place properly.
_ Removing
Perform steps 1 and 2 of the mounting procedure in the reverse order.