25
Version 0.2
SCS-1002
Hospital and refrigerator alarm system
All SCs start operation automatically once the power is supplied.
To power the SCs, firstly, turn on a circuit breaker on the main switchboard, then,
transfer a power button to
ON
position – on SCS-SPS, or press
Power
button – on
SCS-CU, if it is not active.
At least 30 sec after switching on, the System will automatically transfer to wait
mode.
3.2.4
System settings
To provide the System settings, act as follows:
a)
power up the SCs;
b)
follow the steps below to set up RCD (except SCS-WP):
–
set up or reset a password (except SCS-SPS);
– set up a reset timer (20 to 80 sec);
– set up a quantity of controlled calling stations;
– set up a name and background type of the station;
– set up a timer of postponed sound signaling;
– set up an operation mode of reset button (reset or mute);
c)
provide settings of SCS-WP – set up an operation mode of reset button (reset
or mute);
d)
provide settings of controllers – set up a network address using DIP-switches,
see table 13.
Table 13 – Network addresses of controllers
Address
DIP-switch SW2
DIP:5
DIP:4
DIP:3
DIP:2
DIP:1
0
0
0
0
0
0
1
0
0
0
0
1
2
0
0
0
1
0
3
0
0
0
1
1
4
0
0
1
0
0
5
0
0
1
0
1
6
0
0
1
1
0
7
0
0
1
1
1
8
0
1
0
0
0
9
0
1
0
0
1
10
0
1
0
1
0
Содержание SCS-1002
Страница 40: ...Version 0 2 40 SCS 1002 Operating manual Figure А 5 Connection of several SCS SPS 5 by means of SCS KP 6 ...
Страница 50: ...Version 0 2 50 SCS 1002 Operating manual SCS CU PM SCS CU WM ...
Страница 52: ...Version 0 2 52 SCS 1002 Operating manual SCS CU REP PM SCS CU REP WM ...
Страница 60: ...Version 0 2 60 SCS 1002 Operating manual SCS BPC wall mounting w casing SCS BPC panel mounting w o casing ...
Страница 65: ...65 Version 0 2 SCS 1002 Hospital and refrigerator alarm system SCS KP 6 IP22 type SCS KP 6 IP56 type ...
Страница 80: ...Version 0 2 80 SCS 1002 Operating manual FOR NOTES ...