31
Version 0.2
SCS-1002
Hospital and refrigerator alarm system
Table 16 – CL № 1. Exterior inspection of the SC
To be done
Routine
Man-hours per
1 SC
Visually examine the
SC
Check exterior of the SC, mechanical damage; paint defects
must be absent; legends are to be read easily.
Clean up the SC surfaces with clean cloth.
Remove severe contamination, parts of corrosion, oil spots
from metal surfaces with soap suds preventing from
penetration inside the SC; clean dry all surfaces with dry
clean cloth and dry up.
Remove contamination from LCD with cloth or tissue
soaked in alcohol. It is prohibited to use hard cloth, paper,
glass cleaners or chemical substance. Do not push hard
against the monitor while cleaning.
If varnish paint coating is damaged, polish it with sand
paper, then clean with alcohol-soaked cloth, cover with
varnish АК-113 and dry up.
1 person
10 minutes
Check cable and
grounding bus
connections
Check that connectors and mounting screw are fastened
tight; provide further fastening if needed.
Check integrity of (no mechanical damage) the cables.
1 person
5 minutes
Table 17 – CL № 2. Function test of the SC
To be done
Routine
Man-hours per
1 SC
Test functionality of
the System
Turn the System’s on.
Boot software – on-screen picture will appear on SCS-CU
and SCS-CU-REP, or green LED
Power
will start glowing
on SCS-SPS and SCS-WP.
1 person
15 minutes
Table 18 – CL № 3. Completeness check, check of SPTA kit and operational
documentation condition
To be done
Routine
Man-hours per 1
SC
Check completeness,
condition of SPTA
kit and operational
documentation
Compare SPTA kit and operational documentation to the list
specified in section “Completeness”.
Check condition of each item, storage terms and make sure
that SPTA kit was re-completed in case of use.
Complete SPTA kit if necessary.
1 person
1 hour
Содержание SCS-1002
Страница 40: ...Version 0 2 40 SCS 1002 Operating manual Figure А 5 Connection of several SCS SPS 5 by means of SCS KP 6 ...
Страница 50: ...Version 0 2 50 SCS 1002 Operating manual SCS CU PM SCS CU WM ...
Страница 52: ...Version 0 2 52 SCS 1002 Operating manual SCS CU REP PM SCS CU REP WM ...
Страница 60: ...Version 0 2 60 SCS 1002 Operating manual SCS BPC wall mounting w casing SCS BPC panel mounting w o casing ...
Страница 65: ...65 Version 0 2 SCS 1002 Hospital and refrigerator alarm system SCS KP 6 IP22 type SCS KP 6 IP56 type ...