TC-10 User Manual
________________________________________________________________________________________________________________
___________________________________________________________________________
version 3.9
page 9
(16) SENSOR B
connector
8-pole connector for temperature SENSOR B.
This connector also provides the POWER OUTPUT (
18
/
19
) for the heating unit
(max. 12V DC and 2.5A) for devices with the DIN connector cable assembly. (e.g.
HPT-2A, see www.alascience.com).
(17) OUTPUT B 10mV/°C
connector
BNC connector monitoring the temperature determined by SENSOR B (
16
),
sensitivity: 10mV / °C.
(18) / (19) POWER OUTPUT Max. 5A DC
connectors
Banana jack connectors for the heating unit. This output supplies a
maximum voltage of 12V DC and a maximum current of 5A.
POWER OUTPUT is also available at the SENSOR A (
13
) and
SENSOR B connectors (
16
) for devices with the DIN cable assembly.
(20) POWER
pressure switch
Switch to turn POWER on (switch pushed) or off (switch released).
(22) Protective earth
/ (21) GROUND
connectors
To avoid ground loops the internal ground of the system is floating, i.e.
it is not connected to the protective earth of the line cable while the
cabinet is always connected via the power supply cable (green / yellow
wire). The system can be grounded by connecting plug (
21
) protective
earth or (
22
) internal ground.
(23) SHUTOFF TEMPERATURE
label
Label indicating the upper SHUTOFF TEMPERATURE (standard is 60 °C).
If the sensor measures the indicated temperature or more the POWER
OUTPUT is disabled.