(11) SENSOR A / PWR OUT switch (
display mode switch)
Switch to select the mode of display A (
2
):
SENSOR A (upper position): the temperature determined by SENSOR A (
13
) is
displayed (XX.X °C)
PWR OUT ( (lower position): the voltage at the POWER OUTPUT (
18
/
19
) is
displayed.
(12) OUTPUT A 10mV/°C
connector
BNC connector monitoring the temperature determined by SENSOR A (
13
),
sensitivity: 10mV / °C.
(13) SENSOR A
connector
8-pole connector for temperature SENSOR A.
This connector also provides the POWER OUTPUT (
18
/
19
) for the heating unit
(max. 12V DC and 5A) for devices with the DIN connector cable assembly. (e.g.
HPC-G, see www.alascience.com).
(14) MODE
switch
Switch to select the operation modes of the TC-10. Four
modes are available:
SENSOR A: the temperature determined by SENSOR
A (
13
) is used for the control system
DIRECT:
the voltage at the POWER OUTPUT (
18
/
19
) can be set directly from the DESIRED
TEMPERATURE potentiometer (
1
) (0-12 V, no control)
EXT. SENSOR (EXTERN SENSOR): an external thermometer connected to BNC (
15
)
(sensitivity 1mV/°C) can be used for the controller loop
SENSOR B: temperature determined by SENSOR B (
16
) is used as feedback temperature in
the controller loop.
(15) EXT. SENSOR 1mV/°C
connector
BNC connector to connect external thermometer with a voltage output of 1mV/°C
(see also (
14
)).