3.5. Modes of Operation 

 
The system can be used in the following modes of operation: 
 
DIRECT MODE: 

output voltage is set directly by the DESIRED TEMP. °C control. 

Internal SENSOR mode:  output voltage is controlled by the PI controller using SENSOR A or 

SENSOR B. 

EXT. SENSOR mode: 

output voltage is controlled by the PI controller using an external 
thermometer. 

 

DIRECT mode 

 

DIRECT mode (no temperature control): the MODE switch (

14

, Figure 1) is set to the DIRECT 

position. The control unit is not active, i.e. the output voltage is preset directly at the DESIRED 
TEMP. control ((

1

, Figure 1). With the setting from 000 to 999 the output voltage increases 

linearly from 0 V to 12 V. The ALARM circuit (see below) is disabled in this mode. 
 
The TC-10 is designed with an unipolar output (0-12 Volt DC). If a resistive load is used, the 
system can be used for heating purposes in all modes (SENSOR and DIRECT). 
 

Internal SENSOR mode 

 
In  this  mode  the  output  voltage  is  controlled  by  the  PI  controller  (PI  [proportional-integral 
controller  system]  activated).  The  temperature  signal  which  is  used  for  the  control  unit  is 
selected by the MODE switch. 

 

The  SENSOR  A  or  SENSOR  B  position  selects  one  of  the  internal  thermometers 
(SENSOR A or SENSOR B) which use a small semiconductor sensor. The measured 
temperature is shown at the respective digital display unit (

2

 or 

6

, Figure 1 respectively, 

XX.X °C). 

 

If no sensor is connected or if the sensor is damaged or short circuited the internal shut-
off unit disconnects the power output. This state is indicated by the red ALARM LED 
(

3

 or 

7

, Figure 1 respectively) located just above the sensor connectors. 

 

Caution

: The temperature sensors are very sensitive to mechanical stress, especially to bending. 

Therefore, they must be handled with great care to avoid breaking of the insulation! 

 

EXT. SENSOR mode 

 
This mode can be used to connect an external thermometer. The calibration of the external input 
(SENS. INPUT BNC, (

15

, Figure 1)) is 1 mV / °C. If this option is used the internal ALARM 

circuit is disabled. 
 
POWER  OUTPUT  is  always  available  at  POWER  OUTPUT,  SENSOR  A  and  SENSOR  B 
connectors. 
 
 

Caution

: Do not operate heating devices with too much voltage! Refer to the respective manual 

of the device for maximum voltage allowed, and use the LIMITER (

8

) when necessary!! 

 
 

Содержание TC-10

Страница 1: ...EM DESCRIPTION FOR THE TC 10 TEMPERATURE CONTROL SYSTEM VERSION 3 9 npi 2018 npi electronic GmbH Bauhofring 16 D 71732 Tamm Germany Phone 49 0 7141 9730230 Fax 49 0 7141 9730240 support npielectronic...

Страница 2: ...s 4 3 TC 10 System 4 3 1 System Description 4 3 2 Description of the Front Panel 5 3 3 Rear Panel Elements 10 3 4 Electronic Control and Protection Circuits 10 3 5 Modes of Operation 11 3 6 Setting th...

Страница 3: ...is to be operated only at 115 230 Volt 60 50 Hz AC Please check for appropriate line voltage before connecting any system to mains Always use a three wire line cord and a mains power plug with a prote...

Страница 4: ...ndard cabinet with a built in power supply and cooling elements for the power devices The TC 10 guarantees low noise operation and has special protection features to prevent the preparation from damag...

Страница 5: ...____________________________________________________________________________________ ___________________________________________________________________________ version 3 9 page 5 3 2 Description of t...

Страница 6: ...working temperature of the controller system 2 TEMPERATURE C A display Digital display to indicate the temperature determined by SENSOR A or the voltage at the POWER OUTPUT connectors 18 19 Switch 11...

Страница 7: ...to indicate that temperature SENSOR B 16 is not connected broken or short circuited The POWER OUTPUT 18 19 shuts down if this LED is on and the MODE switch 14 is set to SENSOR B 8 LIMITER potentiomete...

Страница 8: ...perature SENSOR A This connector also provides the POWER OUTPUT 18 19 for the heating unit max 12V DC and 5A for devices with the DIN connector cable assembly e g HPC G see www alascience com 14 MODE...

Страница 9: ...onnectors for the heating unit This output supplies a maximum voltage of 12V DC and a maximum current of 5A POWER OUTPUT is also available at the SENSOR A 13 and SENSOR B connectors 16 for devices wit...

Страница 10: ...output signal obtained from the thermometer giving an error signal This signal is amplified in the PI controller and transferred to the output stage where it is converted to a high power output signa...

Страница 11: ...de the output voltage is controlled by the PI controller PI proportional integral controller system activated The temperature signal which is used for the control unit is selected by the MODE switch T...

Страница 12: ...ystem the DESIRED TEMPERATURE can be set using the COMMAND INPUT 10mV C connector 10 Figure 1 The COMMAND INPUT is scaled 10mV C i e 370 mV at this connector would set the DESIRED TEMPERATURE to 37 C...

Страница 13: ...__________________________________________________________________ ___________________________________________________________________________ version 3 9 page 13 5 Temperature sensors Block diagram 5...

Страница 14: ...ft with a DIN connector on one end and two 1 mm connectors on the other end The DIN connector is for connection to the SENSOR input at the front panel of the temperature controller The 1 mm connector...

Страница 15: ...__________________________________________________________________________________________ ___________________________________________________________________________ version 3 9 page 15 5 2 Block dia...

Страница 16: ...the instrument Turn all potentiometers to the most left position switch the MODE selector to the position SENSOR A or B depending on which sensor is used as feedback temperature in the controller loop...

Страница 17: ...TION Thus when beginning to use the TC 10 we recommend to try out several first settings of GAIN and INTEGRATION to find out which settings are best for your experimental conditions If you use a perfu...

Страница 18: ...m ALA please follow these hints Always use the LIMITER Do not set the LIMITER to more than 50 for a longer time since the HPT 2A will melt if operated at more than 5V If in doubt measure the output vo...

Страница 19: ...al displays 3 1 2 digits XX X C temperature of SENSOR A or B or XX X V voltage at power output COMMAND INPUT analog input 10 mV C via BNC connector Set value control digital control range up to 60 0 C...

Отзывы: