ELC-01MX User Manual
version 2.6
page 31
Potential output:
BNC connector, range: 10, 100, 1000
Potential low-pass filter:
3-pole BESSEL filter (other options available)
attenuation: -18 dB/octave, corner frequency:
3 kHz
Potential high-pass filter:
1-pole BESSEL filter, other options available,
attenuation: -6 dB/octave, corner frequencies
(Hz): 1, 10, 50, 100, 500
Digital displays:
Display mV/M
Ω
:
3 ½ digits, XXXX mV or XXX M
Ω
Display current:
3 ½ digits, XX.XX nA
Inputs:
Input impedance analog:
100 k
Ω
Input range:
±12 V
Input impedance digital (TTL):
10 k
Ω
Input TTL:
range: 0-5 V, LO: <0.8 V, HI: >1 V
Current stimulus input:
via BNC connectors, sensitivity 10 nA / V
Step gate input:
via BNC connector (TTL)
Gated stimulus:
with ten-turn control of holding current,
resolution: 100 pA, range: ±100 nA
or
with ten-turn control of holding potential,
in VC mode: resolution: 1 mV, range: ±1 V
in VCx10 mode: resolution: 10 mV, range: ±10 V
Polarity:
selectable with toggle switch
Voltage command input:
via BNC connector, sensitivity
÷
10 mV
EPMS-07 SYSTEM
Power Requirements:
115/230 V AC, 60/50 Hz, fuse 2 A / 1A, slow, 45-60 W (depending on the modules plugged
in)
Dimensions:
19” rack-mount cabinet, 3U high (1U = 1 3/4” = 44.45 mm)