5.1 Headstage Elements
•
P
EL
BNC connector for the electrode holder
•
REF
Connector for the reference electrode (differential headstage only)
•
GND
Ground connector
•
TYPE
Type of amplifier, BA
Bridge amplifier, ELC
ELC amplifier
•
R FB
Value of feedback resistor, 10M: 10 M
Ω
, 100 M: 100 M
Ω
•
Headstage cable to amplifier
•
Mounting plate (holding bar or dovetail available on request)
The electrode filled with electrolyte is inserted into an electrode holder (optional) that fits into
the BNC connector of the headstage or into an electrode holder adapter. The electrical
connection between the electrolyte and the headstage is established using a carefully
chlorinated silver wire. Chlorinating of the silver wire is very important since contact of silver
to the electrolyte leads to electrochemical potentials causing varying offset potentials at the
electrode, deterioration of the voltage measurement etc. (for details see Kettenmann and
Grantyn (1992)). An automated chlorinating apparatus (ACl-01) is available from npi for
optimal chlorinating of sliver wires (please contact npi for details).
Ground provides system ground and is linked to the bath via an agar-bridge or a Ag-AgCl
pellet. The headstage is attached to the amplifier with the headstage cable (see Figure 6) and
an 8-pole connector. The headstage is mounted to a insulated mounting plate that fits to most
micromanipulators. A holding bar or dovetail is available on request.
Note: The shield of the BNC connector is linked to the driven shield output and must not be
connected to ground. The headstage enclosure is grounded.
Caution: Please always adhere to the appropriate safety precautions (see chapter 1). Please
turn power off when connecting or disconnecting the headstage from the HEADSTAGE
connector!