Operation Manual
Hardness Tester NOVOTEST T-UD3
Page. 45
,
(2)
where
n
is the number of measurements.
5.
Select the hardness measurement on the user scale
НB.
6.
Select the statistics measurement mode.
7.
Make at least 3 measurements of
𝐻𝐻𝐻𝐻
𝑖𝑖
on hardness reference blocks (excluding explicit
measurement errors).
8.
Determine the basic error of hardness measurement on the HB scale by the formula (2).
9.
If the error exceeds the allowable value (Table 1.2), then it is necessary to calibrate the user
scale according to p. 2.3.7.
10.
Repeat operations 6 through 8.
11.
The basic error of hardness measurement for the HB scale and for the user HB scale should
not exceed ± 10
,0 НRB
.
3.2.6 Determination of the basic error of the instrument on the scale
НV
1.
Select the hardness measurement on the
НV
scale.
2.
Select the statistics measurement mode.
3.
Carry out at least 3 measurements of
𝐻𝐻𝐻𝐻
𝑖𝑖
on hardness reference blocks (excluding explicit
measurement errors).
4.
Determine the basic error of hardness measurement on the НV scale by the formula
:
,
(3)
where
n
is the number of measurements.
5.
Select the hardness measurement on the user scale
НV.
6.
Select the statistics measurement mode.
7.
Make at least 3 measurements of
𝐻𝐻𝐻𝐻
𝑖𝑖
on hardness reference blocks (excluding explicit
measurement errors).
8.
Determine the basic error of hardness measurement on the
НV
scale by the formula (3).
9.
If the error exceeds the allowable value (Table 1.2), then it is necessary to calibrate the user
scale according to p. 2.3.7.
10.
Repeat operations 6 through 8.
11.
The basic error of hardness measurement for the НV scale and
for the user
НV
scale should
not exceed
± 15,0 НV.
3.2.7 Registration of verification results
Positive results of the primary verification of the device are to be mentioned in the SERVICE
AND MAINTENANCE RECORDS.
The results of periodic verification of a device that is recognized as ready for service are
documented by: a protocol of verification in free form, a mark in the device SERVICE AND
MAINTENANCE RECORDS and a device marking.
A negative result is documented by a certificate of the inadequacy of the device, indicating the
cause, and deletion of the previous verification marking.
3.3 Warranty
The following warranty information is valid for all NOVOTEST products.
The manufacturer guarantees the conformity of the device to the requirements of the technical
conditions under the user's compliance with the conditions of transportation, storage and operation,
and timely maintenance at the manufacturer's premises at least once a year.
Содержание T-UD3
Страница 1: ...Combined Hardness Tester NOVOTEST T UD3 Operating Manual 2019...
Страница 42: ...Operation Manual Hardness Tester NOVOTEST T UD3 Page 42 Figure 2 55 Screen saving tab...
Страница 51: ...Operation Manual Hardness Tester NOVOTEST T UD3 Page 51 SERVICE AND MAINTENANCE RECORDS Description Date...
Страница 52: ...Operation Manual Hardness Tester NOVOTEST T UD3 Page 52...