Operation Manual
Hardness Tester NOVOTEST T-UD3
Page. 44
1
2
3
4
5
4. Determination of
the basic error of the
instrument according
to the
НB
scale
3.3.5
Reference
hardness test
blocks
yes
yes
5. Determination of
the basic error of the
instrument according
to the
НV
scale
3.3.6
Reference
hardness test
blocks
yes
yes
3.2.2 Visual inspection
The device standard set should be in accordance with p.1.3.
No foreign objects should be inside the electronic unit, when tilting the device detected by
hearing.
Probes and its parts should not have any corrosion and mechanical damages.
3.2.3 Testing
The device is tested according to p. 1.7, and also:
−
NOVOTEST T-U3 testing is carried out with an Ultrasonic Contact Impedance Probe U1;
−
NOVOTEST T-D3 testing is carried out with a Rebound Leeb Probe.
3.2.4 Determination of the basic error of the instrument according to the HRC scale
1.
Select the HRC hardness measurement.
2.
Select the statistics measurement mode.
3.
Carry out at least 3 measurements of
𝐻𝐻𝐻𝐻𝐻𝐻
𝑖𝑖
on hardness reference blocks (excluding explicit
measurement errors).
4.
Determine the basic error of hardness measurement on the HRC scale using the formula:
,
(1)
where
n
– number of measurements.
5.
Select the hardness measurement on user scale
НRС
.
6.
Select the statistics measurement mode p. 2.3.5.
7.
Make at least 3 measurements of
𝐻𝐻𝐻𝐻𝐻𝐻
𝑖𝑖
on hardness reference blocks (excluding explicit
measurement errors).
8.
Determine the basic error of hardness measurement on the HRC scale by the formula (1).
9.
If the error exceeds the allowable value (Table 1.2), then it is necessary to calibrate the user
scale according to p. 2.3.7.
10.
Repeat operations 6 through 8.
11.
The basic error of hardness measurement for the HRC scale and for the user
НRC
scale should
not exceed ± 2,0 НRC
.
3.2.5 Determination of the basic error of the instrument on the scale HB
1.
Select the hardness measurement on the HB scale.
2.
Select the statistics measurement mode.
3.
Carry out at least 3 measurements of
𝐻𝐻𝐻𝐻
𝑖𝑖
on hardness reference blocks (excluding explicit
measurement errors).
4.
Determine the basic error of hardness measurement on the HB scale by the formula:
Содержание T-UD3
Страница 1: ...Combined Hardness Tester NOVOTEST T UD3 Operating Manual 2019...
Страница 42: ...Operation Manual Hardness Tester NOVOTEST T UD3 Page 42 Figure 2 55 Screen saving tab...
Страница 51: ...Operation Manual Hardness Tester NOVOTEST T UD3 Page 51 SERVICE AND MAINTENANCE RECORDS Description Date...
Страница 52: ...Operation Manual Hardness Tester NOVOTEST T UD3 Page 52...