
Operation Manual
Hardness Tester NOVOTEST T-UD3
Page. 28
2.3.7 Calibration
2.3.7.1. Calibration of scales
To calibrate the probe, 3 samples of material with known hardness value are required. The
hardness range should be wider than the hardness of the material that will be measured in the future
(the value should be a maximum or more, a minimum or less, and an average).
1.
Select <CALIBRATION> from the main menu. Each table cell corresponds to a specific
calibration scale for a particular material (Figure 2.27). All calibrations can be calibrated for
any materials and any scales, and are thus separated to systematize calibration sets.
Figure 2.27 - Calibration Mode
2. Use the navigation keys to select a cell for calibration, for example, HRC for steel (Figure
2.28).
Figure 2.28 - Selecting a cell for calibration
User Scales
Rockwell
C
Brinell
Vickers Ro
с
kwell
Shore
Steel
Alloy steel
Stainless steel
Cast iron
High-strength cast
iron
Aluminum
Brass
Bronze
User Material
Copper
Содержание T-UD3
Страница 1: ...Combined Hardness Tester NOVOTEST T UD3 Operating Manual 2019...
Страница 42: ...Operation Manual Hardness Tester NOVOTEST T UD3 Page 42 Figure 2 55 Screen saving tab...
Страница 51: ...Operation Manual Hardness Tester NOVOTEST T UD3 Page 51 SERVICE AND MAINTENANCE RECORDS Description Date...
Страница 52: ...Operation Manual Hardness Tester NOVOTEST T UD3 Page 52...