Operation Manual
Hardness Tester NOVOTEST T-UD3
Page. 38
6.
After saving the photo using the navigation keys <
>, <
>, <
>, <
> mark
the measurement location on the photo (Figure 2.50). Scale and the value of hardness, as well
as the date and time of measurement (creating a snapshot) are indicated automatically. To
finish the photo fixation and save the measurement in the archive, press <
>.
Note: To save a picture with one measurement point (marker), press <
>. And in order
to mark the next measurement point on the same photo, press the key <
>. If confirm
saving the measurement point (marker) by the key <
> – will be saved photo without the
ability to add a new measurement point (marker).
Figure 2.50 - Marking of the measuring point
2.3.13 Connecting to PC
2.3.13.1 Software setup
When connecting to PC the data may be sent from the device and to the device as well. <Archive>
measurement data are sent from the device. Also the user can save device display pictures on-line.
Color display settings could be load into the device memory (may be changed in settings in the
<PALETTE> section), as well as the archived data may be controlled.
To connect the device to PC do the following:
1.
Copy driver
“DRIVER_FT232RL”
for “Windows 7” and “Windows 8” or
“CP210x_VCP_Windows” for “Windows 10”
(х64 or х32
according to your OS) and the
“AWP UNIVERSAL” program (.zip type archives) to the PC hard drive or download the
updated versions from the official web site.
2.
Unpack the files getting two folders (Figure 2.51): driver and program.
Figure 2.51 – Downloaded software
Содержание T-UD3