CONTROLS AND INDICATORS
PAGE
|
51
TMP (Temperature) LED illuminates green to indicate laser temperature is within limits and
lasing may be enabled; the LED is red, and lasing is disabled if coolant temperature or flow rate is
outside operating limits.
RDY (Ready) LED illuminates yellow when the laser is enabled, indicating that, after a five-
second delay, lasing will begin when a PWM Command signal is applied.
SHT (Shutter) LED illuminates blue to indicate that the electromechanical shutter is Open and
lasing is enabled. The SHT LED is off, and lasing is disabled if the shutter is Closed.
LASE LED illuminates red to indicate the I401 is actively lasing.
Important Note: When mounting optical components to i401 lasers, the 8–32 UNC fasteners
must extend no
further than 6.35 mm (0.25") into the laser’s faceplate.
Figure 3-2 i401 rear panel controls and indicators
5 User I/O Connector – provides a connection point for auxiliary output power, as well as input
and output signals. Refer to the Technical Reference section for pinouts and signal descriptions.
6
Gas Purge Port/Assembly – provides a low-pressure nitrogen (or pure air) connection to
prevent dust and debris from damaging electronic or optical components inside the housing.
7
WATER IN Port – provides a 12 mm inlet connection to i401 cooling system for 12 mm O.D.
cooling tubing.
8
WATER OUT Port – provides a 12 mm outlet connection from i401 cooling system for 12 mm O.D.
cooling tubing.
Содержание i401
Страница 1: ...ENGINEERED BY SYNRAD i401 Laser User Manual...
Страница 43: ...I401 LABEL LOCATIONS PAGE 43 i401 label locations Figure 2 1 Hazard label locations...
Страница 49: ...DECLARATION OF CONFORMITY PAGE 49 Declaration of conformity Figure 2 2 i401 Declaration of conformity...
Страница 97: ...I401 OUTLINE MOUNTING DRAWINGS PAGE 97 Figure 4 29 I401 outline mounting dimensions mounting feet removed...
Страница 98: ...I401 PACKAGING INSTRUCTIONS PAGE 98 i401 packaging instructions Figure 4 30 i401 packaging instructions...
Страница 132: ...INDEX PAGE 11...