
A-14
Prime CCS Instructions for Use Manual
Precision Study Results
Analytical Precision or Repeatability
Three Stat Profile Prime CCS Analyzers were assessed
for within run precision or repeatability within a laboratory
setting. The protocol consisted of within run precision, and
run–to-run precision studies using 3 different quality control
materials.
• Stat Profile Prime Quality Control Level 1
• Stat Profile Prime Quality Control Level 2
• Stat Profile Prime Quality Control Level 3
Within Run Precision Performance
The protocol consisted of 20 replicates per run for each QC
level on each of 3 Stat Profile Prime CCS Analyzers. The
average, SD, CV%, and N for each analyzer for each QC
level and parameter was calculated. The pooled average,
SD, CV%, and N from all 3 analyzers for each QC level and
parameter was calculated.
Содержание Stat Profile PRIME
Страница 1: ...Stat Profile PRIME CCS Analyzer Instructions for Use Manual 52920...
Страница 2: ......
Страница 4: ......
Страница 6: ......
Страница 20: ...1 10 Prime CCS Instructions for Use Manual...
Страница 67: ...A 7 Appendix A Appendix A...
Страница 68: ...A 8 Prime CCS Instructions for Use Manual...
Страница 69: ...A 9 Appendix A Appendix A...
Страница 70: ...A 10 Prime CCS Instructions for Use Manual...
Страница 71: ...A 11 Appendix A Appendix A...
Страница 102: ...B 16 Prime CCS Instructions for Use Manual...