4-4
Prime CCS Instructions for Use Manual
4.2.2 Change Control Cartridge
1. Press the Tool Box
icon.
2. From the System Menu, press the Change Control
Cartridge and wait for pump to stop.
3. Open the door, remove old cartridge.
4 Slide new cartridge in past the front retaining lip.
5. Close door, press the Calibrate
icon.
4.3 Change Sensor Card
1. Press the Tool Box
icon.
2. From the System Menu, press the Change Sensor
Card and wait for pump to stop.
3. Open the door, remove old cartridge.
4 Open Cartridge door. Insert new cartridge. Close
Cartridge door.
5. Close door, press the Calibrate
icon.
NOTE: Touch Sensor Card by the edges and replace it
the same way back as pictured.
Figure 4.4 Sensor Card
Sensor Card
Hold from
Edges