1-9
1. Intr
o.
1 Introduction
1.6.2 Acceptable Anticoagulants
•
Sodium
and
lithium heparins
are the
acceptable
anticoagulants
for use with the analyzer.
• EDTA, citrate, oxalate, and sodium fluoride
ARE
NOT
acceptable for use.
• Depending on the amount of heparin used in the
collection syringe and whether it is filled to capacity
with blood, heparin concentrations of 20 I.U. per mL
to over 100 I.U. per mL may result.
•
Excess
sodium heparin may
elevate
sodium results.
• Liquid or dry heparin when present in
excess
may
cause errors.
Ensure blood collection devices are
filled per manufacturer instructions.
• Our experience suggests that lyophilized lithium
heparin giving a final concentration in blood of not
more than 20 I.U. per mL is acceptable.
CAUTION: Stat Profile Prime CCS Analyzer users
should take careful note of these considerations
when establishing reference intervals and
interpreting results.