
2-8
Prime CCS Instructions for Use Manual
2.3.1 Manual Calibrations
A manually initiated 2-point calibration can be performed
whenever the analyzer displays Ready or Not Ready on
the header bar.
A Not Ready (Not Calibrated) status is displayed after
powering the analyzer on, after replacing some consumable
items or as a result of a system error. When the analyzer
displays Not Ready, samples cannot be run until a 2-point
calibration is performed that successfully calibrates the air
detectors and at least one analyte. To initiate a calibration
from the Not Ready state, press the
Calibrate
icon
on the Menu Bar.
A Ready status indicates the air detectors and one or more
analytes are calibrated and ready for analysis. To manually
calibrate the analyzer from the Ready state press Toolbox
icon
then press
Calibrate
.
Analytes that display an
orange
background may be
uncalibrated. Press the icon and select
Calibrate
if
displayed to initiate a 2-point calibration.
Figure 2.6
Not Ready
Screen and
Ready
Screen
Содержание Stat Profile PRIME
Страница 1: ...Stat Profile PRIME CCS Analyzer Instructions for Use Manual 52920...
Страница 2: ......
Страница 4: ......
Страница 6: ......
Страница 20: ...1 10 Prime CCS Instructions for Use Manual...
Страница 67: ...A 7 Appendix A Appendix A...
Страница 68: ...A 8 Prime CCS Instructions for Use Manual...
Страница 69: ...A 9 Appendix A Appendix A...
Страница 70: ...A 10 Prime CCS Instructions for Use Manual...
Страница 71: ...A 11 Appendix A Appendix A...
Страница 102: ...B 16 Prime CCS Instructions for Use Manual...