5-4
Prime CCS Instructions for Use Manual
5. Pinch together the white probe holder and remove
the probe together with the S-line and air detector
cable and discard.
Figure 5.7 Removing Probe, S-line, and Cable of Air Detector
6. Place new Probe assembly until it clicks into place.
7. Replace the capillary adaptor back onto the probe
8. Reconnect the S-line to the Reference Sensor
9. Reconnect the Air Detector Cable into the analyzer.
10. Close the door.
14. Press the Calibrate
icon.
Pinch
Clamp
S-Line with
Connection to
Reference Sensor
Cable of Air
Detector to
Analyzer