Indications for Use
Indications for Use
The Nonin Model 2500A VET PalmSAT
®
Veterinary Pulse Oximeter with Alarms is indicated for use
in measuring and displaying functional oxygen saturation of arterial hemoglobin (SpO
2
) and pulse
rate for animals. The device is intended for continuous monitoring and/or spot-checking of animals
during both motion and no-motion conditions, and for animals who are well or poorly perfused.
CONTRAINDICATION:
Do not use in an explosive atmosphere or in the presence of flammable anesthetics or gasses.
Warnings
Do not use this device in an MR environment.
This device is not defibrillation proof per IEC 60601-1.
The Model 2500A VET is intended for VETERINARY USE ONLY.
This device is intended only as an adjunct in animal assessment. It must be used in conjunction with other methods of
assessing clinical signs and symptoms.
Inspect the sensor application site at least every 4 hours to ensure correct sensor alignment and tissue integrity. Animal
sensitivity to sensors and/or double-backed adhesive strips may vary due to medical status or skin condition.
Oximeter readings of this device may be affected by the use of an electrosurgical unit (ESU).
Do not use a damaged sensor. If the sensor is damaged in any way, discontinue use immediately and replace the
sensor.
Verify all alarm settings and limits during system startup to ensure that they are set as intended.
A hazard can exist if different presets are used on multiple 2500A VET monitors in one care area.
As with all medical equipment, carefully route cabling to reduce the possibility of animal entanglement, strangulation, or
injury.
This device should not be used adjacent to or stacked with other equipment. If adjacent or stacked use is necessary, the
device should be observed carefully to verify normal operation.
To prevent improper performance and/or injury to the animal, verify compatibility of the monitor, sensor(s), and
accessories before use.
No modifications to this device are allowed as it may affect device performance.
Discontinue use of adhesive tape strips if the animal exhibits an allergic reaction to the adhesive material.
Avoid excessive pressure to the sensor application site as this may cause damage to the skin beneath the sensor.
Because operating environments vary, use caution to ensure that all audible alarms and indicators can be heard. Users
must determine the acceptable audible distance of all alarms.
Do not place this device in an environment where its speaker opening may become blocked; alarms may become
muffled or inaudible.
Turning off the alarm volume creates a situation that is not compliant with relevant safety standards. The alarm silence
indicator is lit solid when the alarm volume is turned off or set below 45 dBA.
When a system fault occurs, the animal will no longer be monitored.
To comply with relevant product safety standards, ensure that all alarm volumes are set appropriately and are audible in
all situations. Do not cover or otherwise hinder any speaker openings.
The device turns off after approximately 10 minutes when at critically low battery capacity.