![Nonin PalmSAT 2500A VET Скачать руководство пользователя страница 27](http://html1.mh-extra.com/html/nonin/palmsat-2500a-vet/palmsat-2500a-vet_operators-manual_1700109027.webp)
21
Detailed Operation
Clearing the Memory
The Memory Clear function deletes the data currently stored in the device’s memory.
NOTE:
If “no” is selected for either
CLr
or
dEL
, Setup mode continues to the calendar and
clock settings.
1. Enter Setup mode, and scroll through the settings until Memory Clear (
CLr
) appears in the
SpO
2
display. Default value is No.
2. Press
Advance
to change value to Yes.
3. Press
On/Off
to accept and move to the next setting (
dEL
). Default value is No.
4. Press
Advance
to change value to
yES
.
5. Press
On/Off
to accept and clear the device memory.
6.
dnE CLr
briefly appears in the SpO
2
and pulse rate displays to show the memory has
cleared.
7. Device exits Setup mode, automatically displays the alarm settings in effect, and begins
normal operation.
Calendar and Clock Settings
NOTE:
Setting the month to “
00
” disables the calendar and clock functions and helps conserve
battery life.
1. Enter Setup mode and scroll through the settings until the calendar year setting (
y
) appears in
the SpO
2
display.
2. Press
Advance
to scroll through the values.
3. Press
On/Off
to accept a value and move to the next setting. After year, select month (
nn
),
day (
d
), hour (
h
), and minute (
nn
).
4. After selecting the last desired setting, press and release
On/Off
.
5. Device exits Setup mode, automatically displays the alarm settings in effect, and begins