20
Detailed Operation
Recalling Alarm Settings
The 2500A VET retains most user settings when the device is turned off. However, if SpO
2
Low
(
02L
) is set below 85% and/or the Alarm Volume (
Adb
) is OFF, these settings revert to their default
values when the device is turned on.
The user may recall the Alarm Volume--OFF setting by using Setup mode to turn on the device and
selecting
rCL yES
. SpO
2
returns to the default if it is set below 85% even when the device is
turned on with Setup mode.
Alarm settings are retained and available for recall for approximately 30 seconds after removing
batteries.
NOTE:
If set below 85%, the SpO
2
Low Alarm setting defaults to 85% every time the unit is
powered off. The SpO
2
Low setting cannot be recalled by turning on the device in Setup mode.
1. Enter Setup mode. Device displays Recall Alarm (
rCL
). Default value is no.
2. Press
Advance
to change value to Yes.
3. Press
On/Off
to accept and recall the previously-adjusted alarm and volume settings.
4. Device exits Setup mode, automatically displays the alarm settings in effect, and begins
normal operation.
NOTE:
Setup mode exits automatically after the Recall Alarms setting is selected.
Reviewing Alarm Settings
At any time during normal operation, alarm limits and volume settings can be reviewed.
1. Press and hold
Advance
for 1 second.
2. All settings individually flash on the display screen.
NOTE:
To stop the alarm review early and return to normal operation, momentarily press
Advance
.
Silencing Audible Alarms
Audible alarms are automatically silenced for the first 2 minutes of normal operation and can be
temporarily silenced during normal operation.
1. Momentarily press
On/Off
to temporarily silence audible alarms (2 minute silence) during
normal operation.
2. Alarm Silence indicator blinks when alarms are temporarily silenced.
3. Press
On/Off
again to cancel the temporary alarm silence.
CAUTION:
Replace batteries within 30 seconds to avoid losing settings (date, time,
and data stored in memory) or corrupting data.
!