![Nonin PalmSAT 2500A VET Скачать руководство пользователя страница 30](http://html1.mh-extra.com/html/nonin/palmsat-2500a-vet/palmsat-2500a-vet_operators-manual_1700109030.webp)
24
Alarms
System Fault Alarms
If the device determines that a system fault exists, an error message (e.g.,
Err E01
) appears in
the SpO
2
and pulse rate displays, along with medium priority alarm indicators. A system fault has
also occurred if the displays and indicators are blank but a continuous audible alarm is sounding.
1. Attempt to clear the error by turning the device off and on.
2. If the problem persists, contact Nonin Technical Service.
Table 5: Medium Priority Alarms
Condition
Visible Indicators
Inadequate signal (i.e., sensor
dislodgement, unusable signal)
Pulse Quality indicator blinks, dash (-) appears in leftmost
digit of SpO
2
display, then SpO
2
and pulse rate displays
freeze for 10 seconds, and then dash appears in middle
digit of SpO
2
and pulse rate displays.
Sensor fault l (i.e., sensor
disconnect, bad cable, Nonin
incompatible sensor)
Pulse Quality indicator is not lit, dash (-) appears in
leftmost digit of SpO
2
display, then SpO
2
and pulse rate
displays freeze for 10 seconds, and then dash appears in
middle digits of SpO
2
and pulse rate displays.
SpO
2
or pulse rate data not
adequate for more than 20 seconds
Dash (-) displays in middle digit of SpO
2
and pulse rate
displays (i.e., out-of-track indication).
Pulse rate data not updated for more
than 30 seconds
Dashes display in pulse rate display.
Battery low
Low Battery indicator flashes.
Battery critical
Low Battery indicator flashes; SpO
2
and pulse rate
displays are blank; and latched Pulse Quality indicator is
red or amber, but not green.
Sound module or system failure
detected
Display error code.
WARNING: When a system fault occurs, the animal will no longer be monitored.