![Nonin Avant 9700 Скачать руководство пользователя страница 44](http://html1.mh-extra.com/html/nonin/avant-9700/avant-9700_operators-manual_1700082044.webp)
38
Technical Information
Equipment Response Time
If the signal from the sensor is inadequate, the last measured SpO
2
and pulse rate values freeze
for 10 seconds and are then replaced with dashes.
Example - SpO
2
Exponential Averaging
SpO
2
decreases 0.75% per second (7.5% over 10 seconds)
Pulse Rate = 75 BPM
Specific to this example:
• The response of the 4-beat average is 1.5 seconds.
• The response of the 8-beat average is 3 seconds.
SpO
2
Values
Average
Latency
Standard/Fast Averaged SpO
2
4 beat exponential
2 beats
Extended Averaged SpO
2
8 beat exponential
2 beats
Pulse Rate Values
Response
Latency
Standard/Fast Averaged Pulse Rate
4 beat exponential
2 beats
Extended Averaged Pulse Rate
8 beat exponential
2 beats
Equipment Delays
Delay
Display Update Delay
1.5 seconds
Alarm Signal Generation Delay
0 seconds