35
Technical Information
Table 10: Electromagnetic Immunity
Immunity Test
IEC 60601
Test Level
Compliance Level
Electromagnetic
Environment—Guidance
This device is intended for use in the electromagnetic environment specified below. The customer and/or
user of this device should ensure that it is used in such an environment.
Electrostatic Discharge
(ESD)
IEC 61000-4-2
±6 kV contact
±8 kV air
±6 kV contact
±8 kV air
Floors should be wood,
concrete, or ceramic tile. If
floors are covered with
synthetic material, relative
humidity should be at least
30%.
Electrical Fast
Transient/Burst
IEC 61000-4-4
±2 kV for power supply
lines
±1 kV for input/output
lines
±2 kV
±1 kV
Mains power quality should be
that of a typical commercial or
hospital environment.
Surge
IEC 61000-4-5
±1 kV differential mode
±2 kV common mode
±1 kV
±2 kV
Mains power quality should be
that of a typical commercial or
hospital environment.
Voltage dips, short
interruptions, and
voltage variations on
power supply input
lines
IEC 61000-4-11
±5% U
T
(>95% dip in
U
T
) for 0.5 cycle
±40% U
T
(60% dip in
U
T
) for 5 cycles
±70% U
T
(30% dip in
U
T
) for 25 cycles
<5% U
T
(>95% dip in
U
T
) for 5 sec.
±5% U
T
(>95% dip in
U
T
) for 0.5 cycle
±40% U
T
(60% dip in
U
T
) for 5 cycles
±70% U
T
(30% dip in
U
T
) for 25 cycles
<5% U
T
(>95% dip in U
T
)
for 5 sec.
Mains power quality should be
that of a typical commercial or
hospital environment.
Power Frequency (50/
60 Hz) Magnetic Field
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical commercial
or hospital environment.
NOTE:
U
T
is the AC mains voltage before application of the test level.