19
Operating the Avant 9700
NOTE:
The Nurse Call feature overrides silenced alarms.
The Nurse Call Circuit Switch, located near the option switch, allows users to select NO
(Normally Open—the left position) or NC (Normally Closed—the right position). The hospital’s
nurse call system configuration will determine whether the Nurse Call Circuit Switch on the
device must be placed in the Normally Open (NO) or Normally Closed (NC).
Figure 6: Monitor Rear View
NOTE:
It is the user’s responsibility to implement the interface between the Nurse Call
system and the Avant 9700, and to adequately test the interface between the Avant 9700 and
the Nurse Call system to ensure operation.
Electrical State of Nurse Call Output
Nurse Call
Switch Position
Nurse Call
Not Active
Nurse Call
Active
Standby
Mode
Right
Open
Closed
Closed
Left
Closed Open
Open
1
2
3
4
5
6
7
8
Nurse Call
Connection
RS232 Serial
Connection
Power
Connection
Nurse Call
Switch