14
Operating the Avant 9700
NOTE:
When option switch 6 is in the DOWN position, the display intensity is automatically
reduced when changing between AC and battery power.
User-Defined Defaults
User-Defined Default functions are restricted to trained users, and they require multiple button
presses in order to prevent accidental activation.
Record an event
marker.
Press
the ON/STANDBY button while the device
is on.
Silence audible
alarms
Press
the Alarm Silence button.
Cancel locked
alarms
Press
the Alarm Silence button.
Display intensity
(brightness)
Hold
the LCD intensity button until the appropriate
brightness is displayed.
Table 4: User-Defined Defaults
Function
Button
Instruction
Recall Previous
User-Defined
Settings
Press
and
hold
the Settings button while
turning on the Avant 9700.
NOTE:
Previous user-defined settings can also
be recalled within the Parameter Adjustment
menu, which is accessed by
pressing
the
Settings button.
Play back and/or
clear patient data
Press
and
hold
the LCD Screens button while
turning on the Avant 9700.
Follow the display prompts after the
PLy bAC
message clears.
Table 3: Basic Functions (Continued)
Function
Button
Instruction
+
+