33
Status panel
•
ON Red – The interlock loop back circuit is shorted to earth or open. For
example, a connected door switch is open.
Status
Indicates the operational state of the laser amplifier.
•
ON Red – An fault in the modules operation has occurred producing an error
message in the status panel. Errors generated are listed below the Status
indicator – see
for a list of possible errors.
•
ON Amber – The laser amplifier is ready.
•
ON Green – The laser amplifier is operating normally and emissions are enabled.
System info
The System Info section shows the following:
•
Laser Serial Number
•
Laser Firmware Revision
Note:
System info is only displayed if the option is checked in the
View
settings - see
.
Measurements
The system temperature is displayed.
Note:
Measurements are only displayed if the option is checked in the View menu -
see
.
Emission button
The
Emission
button turns the laser amplifier emission ON. The button indicator is lit
steady ON RED when any laser emissions are generated. Otherwise, it is OFF
Grey.
Warning:
Class 3b or Class 4 laser emissions are emitted when the
Emission
button
LED is ON RED.
Note:
A BOOSTIK is a laser amplifier, as such its emission can only be enabled when
its connected seed laser emission is enabled. Clicking the
Emission
button without a
seed laser signal at the input will fail and raise error code 60.
CONTROL settings
CONTROL settings are accessible by clicking the gear icon
in the upper right
corner of the
Status Panel
. Clicking the gear icon displays a drop-down menu of
:
Содержание Koheras BOOSTIK
Страница 1: ...Item 800 631 01 Koheras BOOSTIK PRODUCT GUIDE Narrow linewidth laser amplifier...
Страница 10: ...10...
Страница 14: ...14...
Страница 44: ...Module overview 44...
Страница 51: ...51 Graphing Figure 37 Graphing tab X axis setting Y axes settings...
Страница 52: ...Graphing 52...
Страница 58: ...58 2...
Страница 60: ...60 2...
Страница 66: ...66...
Страница 67: ......
Страница 68: ...1 Koheras BOOSTIK Product Description Revision 1 0 09 2021 W 10456...