Error
Possible
Reason
Correction
Implausible
measurement
values
Hydraulically
unsuitable
measurement
place
See Error: “Unstable measurement values”.
External level
signals
Check for correct connection.
Check if cables are crushed.
Check for short circuits and improper resistive
loads or current consumers without galvanic isola-
tion.
Check measurement range and span.
Check input signal in I/O menu.
Sensor
Check for correct connection.
Check if cables are crushed.
Check for extensions/cable types, short circuits,
surge arresters or improper resistive loads.
Check level signal, echo profile, flow velocity sig-
nal, cable parameters and temperature in I/O
menu.
Check if sensor is installed on a vibration-free
place. Check sensor installation (towards flow
direction, horizontal installation), check sensor for
soiling.
Programming
Check if the correct shape of measurement place
has been set, check dimensions (observe units),
sensor type, sensor installation height etc.
Faulty relay output
Connection
Check connections on terminal clamp strip.
Check power supply of external control relays.
In I/O menu check signals to be output.
Check output control function in calibration menu.
Programming
Check if relay outputs are enabled.
Check if outputs are correctly assigned to respec-
tive output channels.
Check additional values such as impulse parame-
ters, limit values, logic etc.
Содержание OCM F
Страница 105: ...Approvals and Certificates OCM F rev 05 06 04 2018 page 105 Approvals and Certificates...
Страница 106: ...Instruction Manual OCM F page 106 OCM F rev 05 06 04 2018...
Страница 107: ...Approvals and Certificates OCM F rev 05 06 04 2018 page 107...
Страница 108: ...Instruction Manual OCM F page 108 OCM F rev 05 06 04 2018...
Страница 109: ...Approvals and Certificates OCM F rev 05 06 04 2018 page 109...
Страница 110: ...Instruction Manual OCM F page 110 OCM F rev 05 06 04 2018...