Instruction Manual
OCM F
page 42
OCM F - rev. 05 / 06.04.2018
21
Operation Mode (RUN)
This menu is a display menu for standard operation mode and indicates day values and error
messages. Containing the following submenus, it is not required for parameter setting:
Submenues
>day values<
and
>error messages<
:
>Day values<
>Info<
-
Selecting the >Info< submenu allows you to
view the total flow values of the past 14 days
(see Fig. 21-1).
Presumption: the transmitter was operated
without any interruption in the past 14 days.
Otherwise it shows the total for the uninter-
rupted days of operation.
-
When selected first the current date and the
first three saved days are visible. Browse to
other days by using the key >down<.
-
The oldest day total will be overwritten as
soon as the 24h-total of the 15th day has
been created (circular memory function).
>Cycle<
-
The flow totals of 24 hours will be indicated.
Totalisation normally is carried out at 00:00 h
(midnight). If desired, this value can be
modified under >Cycle< (see Fig. 21-2).
>Erase counter<
-
All day counters can be erased under
>Erase counter<. Due to safety reasons it is
required to enter the system-PIN and con-
firmation with >ENTER< after erasing.
1
Time of day totalising
2
Day values (and unit)
3
Current day with cumulated total
4
24 h-day totals
5
Date
Fig. 21-1
Display day values / info
Содержание OCM F
Страница 105: ...Approvals and Certificates OCM F rev 05 06 04 2018 page 105 Approvals and Certificates...
Страница 106: ...Instruction Manual OCM F page 106 OCM F rev 05 06 04 2018...
Страница 107: ...Approvals and Certificates OCM F rev 05 06 04 2018 page 107...
Страница 108: ...Instruction Manual OCM F page 108 OCM F rev 05 06 04 2018...
Страница 109: ...Approvals and Certificates OCM F rev 05 06 04 2018 page 109...
Страница 110: ...Instruction Manual OCM F page 110 OCM F rev 05 06 04 2018...