Parameter Setting
OCM F - rev. 05 / 06.04.2018
page 45
Fig. 22-2
System information / info 1…4
>Unit system<
Here the unit systems can be pre-selected.
Available units:
>metric<
l/s, m³/h, cm/s etc.
>UK-english<
ft, in, gal/s etc.
>US-english<
fps, mgd etc.
>Units<
For each of the three measured/calculated values
>Flow rate<
>Veloc.<
>Level<
>Total<
a unit which appears on the display can be selected. Depending on
the unit system selected, there are various units available.
Fig. 22-3
Selecting the units system and units
>Display
format<
Choose the display format (decimal position) to indicate flow rates,
velocity, level and total.
The decimal positions can be selected. Since the transmitter cannot
indicate more than a maximum of five digits (incl. comma/dot), the
decimal places, however, may be reduced as soon as multiple digits
need to be indicated left of the decimal point.
Example:
a total of 10 litres is indicated as 10.00 (xx.yy) instead of x.yyy
Fig. 22-4
Selecting the display format