MICROSCOPE
.
VI Basic Operations
6 Using the Microscope Control
- 146 -
Condenser <Changes condenser module >
Compatible only with the Nikon 90i microscope.
1
)
In the condenser column
, select a condenser number to which your desired condenser
is registered. The condenser will change to the one you select.
2
)
After switching condensers, confirm that the selected number icon is marked with a red
border and that the condenser number and name are displayed.
NOTE:
*
Before using this function, you must set up the condenser information from
a PC with the setup software. For more information, refer to the instruction
manual for your microscope.
*
Right-click on an icon representing a condenser number, and the names of
all registered condensers will be displayed. You also can select the desired
condenser from within this display to change condensers.
Diascopic illumination aperture diaphragm
<Adjusts the diascopic illumination aperture diaphragm>
Compatible only with the Nikon 90i microscope.
1
)
In the diascopic illumination aperture diaphragm column
, select the
icon to open
the diaphragm in increments of 0.1. Or select the
icon to close the diaphragm in
decrements of 0.1.
2
)
Briefly depress the mouse button to adjust the diaphragm in three-increment steps.
NOTE: The level indicator shows the current aperture of the diaphragm in
increments of 1.4 in the range 2.0 to 30.5.
Diascopic illumination field diaphragm
<Adjusts the diascopic illumination field diaphragm>
Compatible only with the Nikon 90i microscope.
1
)
In the diascopic illumination field diaphragm column
, select the
icon to open the
diaphragm in increments of 0.1. Or select the
icon to close the diaphragm in
decrements of 0.1.
2
)
Briefly depress the mouse button to adjust the diaphragm in three-increment steps.