![Nidek Medical LM-7 Скачать руководство пользователя страница 99](http://html1.mh-extra.com/html/nidek-medical/lm-7/lm-7_operators-manual_1688046099.webp)
97
6
SPECIFICATIONS AND TECHNICAL
INFORMATION
6.1
Specifications
Measurement items
• Spherical power (spectacle
lenses)
-25.00 to +25.00 D, 0.01/ 0.06/ 0.12/ 0.25 D increments
Measurement range (D)
Accuracy (D)
< 0
≥
-5
> 0
≤
+5
±0.06
< -5
≥
-10
> +5
≤
+10
±0.09
< -10
≥
-15
> +10
≤
+15
±0.12
< -15
≥
-20
> +15
≤
+20
±0.18
< -20
> +20
±0.25
• Spherical power (contact lenses)
-25.00 to +25.00 D (BC = 6.00 to 9.00), 0.01/ 0.06/ 0.12/ 0.25 D incre-
ments
• Cylindrical power
0.00 to ±10.00 D (-, +/-, +), 0.01/ 0.06/ 0.12/ 0.25 D increments
• Cylinder axis
0° to 180°, 1° increments
• Addition power
+0.4 to +10.00 D (first add, second add), 0.01/ 0.06/ 0.12/ 0.25 D
increments
• Prism power
0.00 to 20.00
∆
, 0.01/ 0.06/ 0.12/ 0.25
∆
increments
Measurement range (
∆
)
Accuracy (
∆
)
> 0
≤
5
0.1
> 5
≤
10
0.2
> 10
≤
15
0.3
> 15
≤
20
0.4
• Prism measurement
Polar coordinates (
∆
,
θ
), Rectangular coordinates (Base in/out, Base
up/down)
• UV transmittance
0 to 100%: Transmittance of central wavelength 365 nm (UV-A). 1%/
5% increments
• Light source wavelength
538 nm (green), 365 nm (during UV transmittance measurement)
1707141668223 承認済
Uncontrolled copy when printed.
Содержание LM-7
Страница 1: ...LM 7 LM 7P AUTO LENSMETER OPERATOR S MANUAL 1707141668223 Uncontrolled copy when printed...
Страница 6: ...4 1707141668223 Uncontrolled copy when printed...
Страница 12: ...10 1 SAFETY PRECAUTIONS Labels and Symbols 1707141668223 Uncontrolled copy when printed...
Страница 88: ...86 4 CONNECTION AND SETTINGS Parameter Settings 1707141668223 Uncontrolled copy when printed...