43
3. OPERATING PROCEDURE : Progressive Power Lens Measurement
2
Set a lens.
❖
The following is the procedure for measuring the right-eye lens.
3
Measure the distance power.
1)
Bring the intermediate portion (progressive zone)
onto the nosepiece.
Place a portion that is slightly lower than the center of
the lens on the nosepiece.
The target
indicating the distance portion is dis-
played.
The arrow shows the direction in which the lens
should be moved.
2)
Move the lens horizontally to align the target to the
vertical line of the crossline.
3)
Slowly move the lens backward (upward on the
screen) to align the target to the crossline.
If the target is shifted to the right or left from the verti-
cal line of the crossline while the lens is moved, move
the lens horizontally.
When the vertical position is roughly aligned, the
arrow
disappears.
4)
Precisely adjust the lens vertically and horizontally to
align the target to the crossline.
When the target changes to
(cross) and the
measured data stabilizes, the measured data is auto-
matically locked in.
The screen automatically changes to add power mea-
surement for the near portion.
• To correctly measure the progressive power lens, note the following.
• First align the lens horizontally, then vertically.
• Move the lens while the frames are constantly in contact with the lens table.
• Ensure that the lens back surface is constantly in contact with the nosepiece.
1707141668223 承認済
Uncontrolled copy when printed.
Содержание LM-7
Страница 1: ...LM 7 LM 7P AUTO LENSMETER OPERATOR S MANUAL 1707141668223 Uncontrolled copy when printed...
Страница 6: ...4 1707141668223 Uncontrolled copy when printed...
Страница 12: ...10 1 SAFETY PRECAUTIONS Labels and Symbols 1707141668223 Uncontrolled copy when printed...
Страница 88: ...86 4 CONNECTION AND SETTINGS Parameter Settings 1707141668223 Uncontrolled copy when printed...