3. OPERATING PROCEDURE : Single Vision Lens Measurement
●
Locking in measured data with the auto read function
When the Auto read S or Auto read R/L parameter is set to “
(cross)” or “
(large cross)”,
measured data is automatically locked in after the target is aligned as well as when the read but-
ton is pressed.
To retry measurement, align the target and press the read button again.
4
Measure the other lens if necessary (in the R/L state).
Switch the R/L measurement select button to
or
.
Follow the same steps as the first lens.
5
Press the print button
to print the measured results (or perform data transmis-
sion or writing to the Eye Care card).
For details of printing, see
.
Parameter
Auto read function
Auto read S
Single state
Auto read R/L
R/L state
• When the Auto read S or Auto read R/L parameter is set to “
(cross)”, auto read is performed
after the target changes to
(cross). If alignment is performed until the target changes to
(large cross), auto read is performed again.
• When the Auto R/L parameter is set to “On”, R and L are changed automatically after the measured
lens is removed.
• When the Auto print S or Auto print R/L parameter is set to “On”, the measured results are printed
automatically after the measured lens is removed.
1707141668223 承認済
Uncontrolled copy when printed.
Содержание LM-7
Страница 1: ...LM 7 LM 7P AUTO LENSMETER OPERATOR S MANUAL 1707141668223 Uncontrolled copy when printed...
Страница 6: ...4 1707141668223 Uncontrolled copy when printed...
Страница 12: ...10 1 SAFETY PRECAUTIONS Labels and Symbols 1707141668223 Uncontrolled copy when printed...
Страница 88: ...86 4 CONNECTION AND SETTINGS Parameter Settings 1707141668223 Uncontrolled copy when printed...