70
4. CONNECTION AND SETTINGS : Operation when Peripheral Devices are Connected
4.1.2
Connecting AR/ARK/RKT
Measured data of the LM-7/LM-7P can be printed with the printer of the connected AR, ARK, or RKT.
Connect the NIDEK AR, ARK, or RKT to the communication port of the LM-7/LM-7P and set the
parameters.
●
Parameter setting procedure
1)
Set the communication related parameters of the LM-7/LM-7P.
For the setting procedure, see
“4.2 Parameter Settings” (page 75)
.
2)
Set the parameters of the connected device.
Connectable equipment
AR-600/330A series, ARK-700/530A series, AR-1/ARK-1 series, RKT-7700/
TONOREF II/TONOREF III
Parameter
Setting contents
Printer
AR print
Com mode
NIDEK
Baud rate
9600
Parity
Odd
Data bits
8bit
Stop bits
1bit
Connecting device
Parameter
Setting contents
AR/ARK series
IN port (LM)
NIDEK
LM Data Prt.
YES
Baud-Rate
9600
Bit Length
8
RKT-7700/TONOREF II/TON-
OREF III
BAUD-RATE
9600
BIT LENGTH
8
LM DATA PRINT
YES
• For the setting procedure, refer to the operator's manual for the connected device.
• When printing with the AR, ARK, or RKT, the font is different from that printed with the LM-7/LM-
7P.
• The Print density, Paper cut, Auto cutter, Economy print, QR code, and PD layout parameters are
not supported.
For PD, only a value is printed, and spacing between lines is unchanged.
1707141668223 承認済
Uncontrolled copy when printed.
Содержание LM-7
Страница 1: ...LM 7 LM 7P AUTO LENSMETER OPERATOR S MANUAL 1707141668223 Uncontrolled copy when printed...
Страница 6: ...4 1707141668223 Uncontrolled copy when printed...
Страница 12: ...10 1 SAFETY PRECAUTIONS Labels and Symbols 1707141668223 Uncontrolled copy when printed...
Страница 88: ...86 4 CONNECTION AND SETTINGS Parameter Settings 1707141668223 Uncontrolled copy when printed...