4. CONNECTION AND SETTINGS : Parameter Settings
❖
The following parameters are displayed only when “LAN” or “WLAN” is selected.
49
Prism Tx
Off, On,
Display
Turns on or off the prism ON/OFF button.
Off: The prism ON/OFF button is normally turned off and measured prism data is not trans-
mitted.
On: The prism ON/OFF button is normally turned on and measured prism data is transmitted.
Display: When measured prism data is displayed on the screen, it is transmitted. When data
is not displayed, it is not transmitted. Display of measured prism data is toggled with the
prism ON/OFF button in the measured results display area.
Even when the setting is “On” or “Off”, whether or not to transmit measured prism data can be
toggled with the prism ON/OFF button.
50
Source
1
Sets the port number for the source (LM-7/LM-7P).
Pressing the item displays the numeric keypad. Enter a value (0 to 255). Enter different num-
bers for “Source” and “Destination”.
51
Destination
0
Sets the port number for the destination (receiving device).
Pressing the item displays the numeric keypad. Enter a value (0 to 255). Enter different num-
bers for “Source” and “Destination”.
52
Request
2.0(second)
Sets the response waiting time (time-out) for command requests.
Pressing the item displays the numeric keypad. Enter a value (0.1 to 10.0).
53
Reception
10(second)
Sets the waiting time (time-out) for receiving commands.
Pressing the item displays the numeric keypad. Enter a value (1 to 100).
54
Response
2.0(second)
Sets the waiting time (time-out) for responding commands.
Pressing the item displays the numeric keypad. Enter a value (0.1 to 10.0).
55
Re-request
2.0(second)
Sets the response waiting time (time-out) for command requests.
Pressing the item displays the numeric keypad. Enter a value (0.1 to 10.0).
56
Retry
3
Sets the number of retries when communication cannot be established.
Pressing the item displays the numeric keypad. Enter a value (0 to 5).
57
LAN
Off, LAN, WLAN
Selects from the detected LAN communication function (undetected function is not displayed).
For the LAN communication, the optional LAN board or WLAN module needs to be mounted.
Parameter
Setting contents
58
MAC address
ex.)
FF-FF-FF-FF-FF-FF
Displays the MAC address.
The setting cannot be changed on the parameter screen.
59
File format
NIDEK_V1.00, NIDEK_V1.01
Selects the file format.
Parameter
Setting contents
1707141668223 承認済
Uncontrolled copy when printed.
Содержание LM-7
Страница 1: ...LM 7 LM 7P AUTO LENSMETER OPERATOR S MANUAL 1707141668223 Uncontrolled copy when printed...
Страница 6: ...4 1707141668223 Uncontrolled copy when printed...
Страница 12: ...10 1 SAFETY PRECAUTIONS Labels and Symbols 1707141668223 Uncontrolled copy when printed...
Страница 88: ...86 4 CONNECTION AND SETTINGS Parameter Settings 1707141668223 Uncontrolled copy when printed...