5. MAINTENANCE : Touch Screen Calibration
5.3
Touch Screen Calibration
If the position touched on the screen and the response position do not match, perform the touch
screen calibration.
1
While pressing the read button, turn on
the
power switch.
Continue pressing the read button until the calibration
screen is displayed.
2
Press the blue squares that appear on the
screen in order.
When the squares in all four corners are pressed, the cali-
bration is complete.
• Pressing with a rod with a narrow, rounded tip such as a touch pen allows more accurate calibration.