CHECK AND ADJUSTMENT PROCEDURES: Groove Depth Check and Adjustment
2)
Press the numeric field of “Dep:” in the “Groove
size” field.
The numeric keypad appears.
3)
Enter a value with the numeric keypad and
press .
Enter 0.0 to “Dep:”.
Enter 0.6 (minimum value) to “Wid:”.
6
Press .
The layout screen appears.
7
Refer to the following procedure for the processing method after size setting on the
layout screen.
8
Check the groove depth of the processed lens.
•
Check that the groove is slightly made (the depth is within 0.10 mm) on the processed lens.
9
When the groove cannot be found or is too deep, change the “Groove depth” param-
eter.