31
CHECK AND ADJUSTMENT PROCEDURES: Size Check and Adjustment
3.5.2
Size adjustment
1
Enter service mode from the LE-700 home screen.
“3.2.1 Entering service mode” (page 20)
2
Press the “Grinding” tab.
3
Press Tab 1.
4
Press the numeric field of “Master size
adjustment”.
The numeric field turns orange and the numeric key-
pad appears.
• When the outer diameter is out of the range, perform the surface measurement calibration a single
time.
“3.4 Surface Measurement (LMU) Calibration” (page 24)
• If the outer diameter is out of the range even after the surface measurement calibration, adjust the
size in the following procedure.
Содержание LE-700
Страница 6: ...IV ...
Страница 24: ...18 INSTALLATION Cooling Water and Cleaning Water ...
Страница 78: ...72 CHECK AND ADJUSTMENT PROCEDURES Date and Time Setting ...