45
CHECK AND ADJUSTMENT PROCEDURES: Size Check and Adjustment
6
Press Circle 45
.
The processing condition screen appears.
7
Specify each processing condition.
1)
Set each item as follows:
•
For flat edge polishing
“Material”
→
CR39
“Type”
→
Single
“Frame”
→
Flat
“Soft”
→
None
“Polish”
→
Exec
“Layout”
→
Passive
“FC”
→
None
•
For bevel polishing
“Material”
→
CR39
“Type”
→
Single
“Frame”
→
Metal
“Bevel position”
→
Auto
“Soft”
→
None
“Polish”
→
Exec
“Layout”
→
Passive
“FC”
→
None
2)
Press .
The layout screen appears.
8
Continue the processing method on the layout screen using the same procedures
explained by the following steps starting from size setting.
to
9
Confirm that the lens edge is not left unpolished or burnt.
“3.8.1 Polishing check” (page 42)
10
until no polished or burnt portions are left.