CHECK AND ADJUSTMENT PROCEDURES: Polished Bevel Position Check and Adjustment
3.9
Polished Bevel Position Check and Adjustment
Check the edge surface of a plastic lens bevel polished
using the Circle 45 internal data. If one side of the bev-
eled surfaces is left unpolished, change the Bevel posi-
tion (Polish) parameter.
3.9.1
Polished bevel position check
1
Enter service mode from the LE-700 home screen.
“3.2.1 Entering service mode” (page 20)
2
Press the “Maintenance” tab.
The maintenance screen appears.
3
Press Circle 45
.
The processing condition screen appears.
4
Specify the bevel polishing conditions.
1)
Set each item as follows:
“Material”
→
CR39
“Type”
→
Single
“Frame”
→
Metal
“Bevel position”
→
Auto
“Soft”
→
None
“Polish”
→
Exec
“Layout”
→
Passive
“FC”
→
None
2)
Press .
The layout screen appears.
5
Continue the processing method on the layout screen using the same procedures
explained by the following steps starting from size setting.
to
Bevel surfaces