:INPut[1|2]:VOLTage:OVERload[:LEVel] <over_level>
:INPut[1|2]:VOLTage:OVERload[:LEVel]?
Description:
Sets/queries the excessive input detection level.
Command prefix: with CH1: INPut1, with CH2: INPut2.
When the channel specification is abbreviated to INPut: the
value is set to CH1.
Parameters:
<
over_level
>
detection level, format
<
NR2
>
, range 0.01 to 19.99
[unit Vrms]
Settings example:
INP1:VOLT:OVER 3.00
sets the excessive input detection level for CH1 to 3.00Vrms
Response:
<
over_level
>
Query example:
INP1:VOLT:OVER?
Response example:
3.00
:INPut:VOLTage:OVERload:RESPonse <over_response>
:INPut:VOLTage:OVERload:RESPonse?
Description:
Sets/queries the process when excessive input is detected.
Parameters:
<
over_response
>
response
format
<
NR1
>
,
range 0/1/2/3 0: Lamp, 1: Beep. 2: Hold, 3: Off
Settings example:
INP:VOLT:OVER:RESP 1
Response:
<
over_response
>
Query example:
INP:VOLT:OVER:RESP?
Response example:
1
Содержание FRA5022
Страница 1: ...NF Corporation FREQUENCY RESPONSE ANALYZER FRA5022 Instruction Manual...
Страница 2: ......
Страница 3: ...FREQUENCY RESPONSE ANALYZER FRA5022 Instruction Manual DA00016981 005...
Страница 26: ...2 2 Installation FRA5022 2 8 Figure 2 5 Removal of feet...
Страница 30: ...FRA5022 2 12...
Страница 92: ...5 5 Command Explanation FRA5022 5 14 Query example SRE Response example 128...
Страница 119: ...FRA5022 5 41...
Страница 131: ...FRA5022 7 8...
Страница 141: ......
Страница 143: ......
Страница 144: ......